Tube name | 8 RXN | 24 RXN | 96 RXN | |
End Preparation | End Prep Buffer | 70 μl | 168 μl | 700 μl |
End Prep Enzymes | 30 μl | 78 μl | 300 μl | |
Damaged DNA Repair Enzymes | 30 μl | 78 μl | 300 μl | |
Adaptor Ligation | Ligation Buffer | 300 μl | 780 μl | 3 ml |
Ligase Enzymes | 100 μl | 240 μl | 1000 μl | |
Amplification | 2X PCR Master Mix | 250 μl | 650 μl | 2.5 mL |
10X PCR Primers | 50 μl | 130 μl | 500 μl |
样本编号 | 样本形态 | 投入量ng | PCR前浓度(ng/mL) | 循环数 | 文库产量 (ng) |
1 | FFPE | 400 | 10.8 | 5 | 1203 |
2 | FFPE | 50 | 1.43 | 8 | 1506 |
3 | FFPE | 400 | 10.8 | 5 | 1392 |
4 | FFPE | 50 | 1.34 | 8 | 1162 |
5 | 新鲜组织 | 400 | 8.06 | 4 | 1518 |
6 | 新鲜组织 | 50 | 0.976 | 7 | 1500 |
分别使用不同起始量的FFPE和新鲜组织样本测试Rapid Plus DNA Lib Prep Kit for Illumina建库效果,扩增循环数为5-8 个,结果表明ABclonal Rapid Plus DNA Lib Prep kit建库结果满足上机需求。
样本编号 | 建库产品 | Yield (ng) | Total Reads | Total Mapped Reads | Duplication Rate | On Target Rate | Insert Size Mean | Mean Coverage | Total SNPs | SNPs Overlap | Private SNPs |
Sample 1 | Hyper | 565 | 10735908 | 10735393 | 35.18 | 81.42 | 150.36 | 871.81 | 155 | 155 | 0 |
Rapid Plus | 750 | 11618662 | 11618229 | 35.62 | 86.20 | 150.02 | 1016.58 | 155 | 155 | 0 | |
Sample 2 | Hyper | 640 | 10224070 | 10223651 | 31.80 | 81.45 | 158.02 | 917.26 | 157 | 157 | 0 |
Rapid Plus | 765 | 14456368 | 14455971 | 36.88 | 85.86 | 155.77 | 1275.83 | 157 | 157 | 0 | |
Sample 3 | Hyper | 785 | 4411574 | 4411273 | 20.56 | 65.03 | 145.72 | 323.24 | 154 | 154 | 0 |
Rapid Plus | 770 | 5918920 | 5918570 | 24.70 | 76.19 | 144.33 | 495.79 | 154 | 154 | 0 |
针对FFPE样品中SNP检测,结果表明Rapid Plus 与 Kapa Hyper 在 SNP位点分析结果上表现出高度一致。